Hygiene checklist for dental practices

  • Each entry in the hygiene record names the person and the time; a tick taken back stays visible with its reason.
  • A closed month carries a checksum that can be re-verified at any time.
  • Hygiene record entries are kept for at least ten years, even after the contract ends.

Nadia Kunz · 07:35Flush waterlines (2 min)

Lena Frei · 07:50Autoclave test run (Helix/Bowie-Dick)

MissedMedicine fridge temperature

The hygiene tasks many Swiss practices do every day, week and month — from flushing the waterlines to the autoclave’s test run. Print it for the steri room, or keep it as a list the team ticks off in Praxovi.

This list is a starting point, not a rule. What applies is set by the Swissmedic guide to reprocessing medical devices, KIGAP, SSO’s quality guidelines on practice hygiene and your canton. Frequencies and device tests follow the manufacturers’ instructions.

Before the first patient
Flush waterlines (2 min)Every unit: turbine, handpiece and 3-way syringe, 2 minutes each, before the first patient.
Autoclave test run (Helix/Bowie-Dick)Before the first load. Photograph the indicator strip or printout.
Medicine fridge temperatureRead the thermometer and type the value. Outside 2–8 °C: note what you did.
Between patients
Sterilizer: record batchBatch number and release per run.
After the last patient
Flush suction systemSuction cleaner per unit, after the last patient.
Treatment rooms: disinfect surfaces
Once a week, on chosen days
Autoclave: vacuum testOnce a week with the chamber empty, per the manufacturer, before the first load. Photograph the printout.
Water quality: unit disinfectionWeekly intensive disinfection of the unit water system per the manufacturer.
Check amalgam separatorFill level indicator; note the date in the maintenance log.
Sharps and amalgam waste containersReplace when three-quarters full; seal and label.
Once a month
Emergency kit: expiry dates and oxygenNote what was replaced.
Autoclave: monthly biological indicatorNot required by KIGAP; the practice decides. Photograph the result.
X-ray: CBCT constancy test(CBCT only)Monthly, on the CBCT. Photograph the test image.
Once a year
X-ray: 2D constancy test (intraoral, OPG)Yearly, on every 2D unit with its image receptor. Photograph the test image or the report.

The same list in Praxovi

Every task that is part of the hygiene record, with its month at a glance. Nothing here can be quietly erased: each entry names the person, the time and — where recorded — the device and lot number; a tick taken back stays visible with its reason. A closed month carries a checksum that can be re-verified at any time, and entries are kept for ten years.

Frequently asked questions

Does the test strip have to be glued into a book?
No. According to the Swissmedic guide, it doesn’t have to be. What is recorded is the result with date, device, lot and person — on paper or in Praxovi.
Can I change the list?
Yes. On paper, strike or add rows. In Praxovi every task is your own template: the time, how often, who does it and how it is confirmed.